Home  
  Products  
  Literature/Applications  
  Service/Support  
  Search  
search
Tosoh: The Chemistry of Innovation
Reversed Phase Columns: Silica-based
8nm Monomeric RP Silica
TSKgel CN-80Ts
TSKgel Octyl-80Ts
TSKgel ODS-80TM
TSKgel ODS-80Ts
TSKgel ODS-80Ts QA
10nm Monomeric RP Silica
2µm Super Series RP Silica
Specialty RP Silica
High Throughput RP Silica


TSKgel ODS-80Ts QA columns were developed specifically for use by QA/QC departments that require highly reproducible separations. These columns are prepared from the same endcapped C18 packing material as TSKgel ODS-80Ts columns, but with narrower manufacturing specifications to meet the demand for high reproducibility.

The variation between different lots of TSKgel ODS-80Ts QA packing material is minimized by selecting batches of TSKgel ODS-80Ts that fall within a very narrow range of specifications, as demonstrated in the table below. In addition, each column must pass demanding specifications for efficiency (N) and peak asymmetry, as are spelled out in the Operating Conditions and Specifications sheet.  The end result is TSKgel ODS-80Ts QA columns that exhibit an unparalleled level of reproducibility for retention, selectivity (k´), efficiency (N), and peak symmetry.


Product Specifications:

 Attribute

Specification Range

Lot-to-Lot Reproducibility (CV%)

 Particle size:
    - Distribution (dp90/dp10)

4.95 - 5.35
1.55 - 1.70

0.6
1.8

 Surface area (m2/g)

410 - 440

0.5

 Pore size (mL/g) silica:

9.0 - 10.0

0.5

 Pore volume (nm):

0.96 - 1.04

0.7

 Carbon content (wt%):

14.0 - 15.0

N/A

 C18 coverage (µmol/m2):

1.71 - 1.99
1.3

 Metal ion content (ppm)
    - Na
    - Al
    - Fe
    - Ti


<10
<10
<10
<10


N/A
N/A
N/A
N/A



Column classification:

TSKgel ODS-80Ts QA columns were submitted to several characterization tests to determine the level of hydrophobic retention, steric selectivity and retention of basic compounds. The results of the characterization tests were used to establish specifications listed in the Table below.

 Parameter

Specification

CV (%)

Test Conditions

 k' naphthalene (hydrophobicity)

1.53 - 1.63

1.3

1

 α triphenylene/o-terphenyl (steric selectivity)

1.21 - 1.25

0.4

1

 k' procainamide (basic compounds)

1.35 - 1.55

2.6

2

 k' phenol

9.25 - 9.85

0.9

2

 k' oxine copper (inertness to chelating compounds)

1.13 - 1.35

3.5

3
 
1. 80% CH3OH/20% H2O, T=40°C
2. 10% ACN/90% 20mmol/L Na2HPO4 (pH6.6, adjusted with 20mmol/L NaH2PO4, T=40°C)
3. 7% ACN/93% 20mmol/L H3PO4, T=40°C


 

Application

To demonstrate the lot-to-lot reproducibility that can be expected when using TSKgel ODS-80Ts QA columns, the contents of an over-the-counter analgesic pain reliever was analyzed on columns from three different lots of packing material. The results are shown in the Figure to the right and Table below. Excellent reproducibility of retention, peak height, peak area, efficiency and peak shape is evident for all three ingredients.

Highly reproducible results can be achieved using the TSKgel ODS-80Ts QA columns as shown in this lot-to-lot reproducibility test.

 Analysis of an over-the-counter analgesic
pain reliever



 

Results

Compound

Lot#
RT (min)

Peak Area
(mv x sec)

Peak Height

N
AF

 caffeine

01W
02W
03W

3.047
3.080
3.080

4.51575 x 102
4.50235 x 102
4.44175 x 102

99.26
94.04
95.63

 10,396
9,597
10,143

1.10
1.13
1.13

 salicylamide

01W
02W
03W

6.190
6.253
6.250

3.96172 x 102
3.94822 x 102
3.90617 x 102

54.06
52.23
52.40

16,441
15,797
16,333
1.03
1.07
1.06

 acetylsalicylic acid

01W
02W
03W

 9.983
10.080
10.063

4.34473 x 102
4.33835 x 102
4.27633 x 102

37.35
36.48
36.28

16,742
16,297
16,649

1.03
1.05
1.05

Literature:

Brochure -- Reversed Phase and HILIC TSKgel HPLC Columns

Operating Conditions and Specifications


 


P/N
Description
Particle
size
Housing material ID
(mm)
Length
(cm)
0018768
TSKgel ODS-80Ts QA
5μm
Stainless Steel 2 15
0018769
TSKgel ODS-80Ts QA
5μm
Stainless Steel 2 25
0018519
TSKgel ODS-80Ts QA
5μm
Stainless Steel 4.6 7.5
0018520
TSKgel ODS-80Ts QA
5μm
Stainless Steel 4.6 15
0018521
TSKgel ODS-80Ts QA
5μm
Stainless Steel 4.6 25
0019325
Guard Cartridge for 2mm ID columns, 3 pk
5μm
Stainless Steel 2 1
0019308
Guard Cartridge Holder for 2mm ID cartridges
 
Stainless Steel 2 1
0019011
Guard Cartridge for 4.6mm ID columns, 3 pk
5μm
Stainless Steel 3.2 1.5
0019018
Guard Cartridge Holder for 4.6mm ID cartridges
 
Stainless Steel 3.2 1.5

 


 

 

Frequently asked questions -- RPC

Return to RP main page

 

Search the Tosoh HPLC Application Database

Tosoh Bioscience Japan, Shiba, Minato-Ku, Tokyo 105-8623, Japan TEL:+81-3-5427-5118 FAX:+81-3-5427-5198 Email:hlc@tosoh.co.jp