Home  
  Products  
  Literature/Applications  
  Service/Support  
  Search  
search
Tosoh: The Chemistry of Innovation
Gel Filtration Chromatography
SW
TSKgel G2000SW
TSKgel G3000SW
TSKgel G4000SW
SWXL
SuperSW

TSKgel SW columns, introduced in 1977, were the first of a long line of high performance Gel Filtration columns that have become synonymous with analyzing protein molecular weights in the emerging field of biotechnology.

Particle Characteristics
SW-type columns are based on highly porous silica particles, the surface of which has been shielded from interacting with proteins by derivatization with ligands containing diol functional groups. TSKgel SW-type columns stand out from other silica- or polymer-based high performance size exclusion columns by virtue of their large pore volumes.

Calibration Curves
The figure below shows the calibration curves for globular proteins, polyethylene oxides and dextrans for each of the three TSKgel SW-type columns. The following table summarizes the molecular weight ranges for G2000SW, G3000SW and G4000SW columns for globular proteins, polyethylene oxides and dextrans that can, at least partially, access the particle pores. For comparative purposes, similar data is provided as well for TSKgel SWXL and SuperSW-type columns.


Polyethylene oxide, dextran and protein calibration curves for TSKgel SW columns




 

Properties and separation ranges for TSKgel SW-type packings (Molecular Weight in Da)

TSKgel packing

Particle
size 
(µm)

Pore
size (nm)

Globular
proteins

Dextrans

Polyethylene
glycols and oxides

SuperSW2000

4

12.5

5,000 - 150,000

1,000 - 30,000

500 - 15,000

G2000SWXL/
BioAssist G2SWXL

5

12.5

5,000 - 150,000

1,000 - 30,000

500 - 15,000

G2000SW

10, 13

12.5

5,000 - 150,000

1,000 - 30,000

500 - 15,000

SuperSW3000

4

25

10,000 - 500,000

2,000 - 70,000

1,000 - 35,000

G3000SWXL/
BioAssist G3SWXL

5

25

10,000 - 500,000

2,000 - 70,000

1,000 - 35,000

G3000SW

10

25

10,000 - 500,000

2,000 - 70,000

1,000 - 35,000

           

G4000SWXL/
BioAssist G4SWXL

8

45

20,000 -7,000,000

4,000 - 500,000

2,000 - 250,000

G4000SW

13, 17

45

20,000 -7,000,000

4,000 - 500,000

2,000 - 250,000

Data generated using the following conditions:

Columns:

Two 4µm, 4.6mmIDx30cm TSKgel SuperSW columns in series; two 5µm, 7.8mmIDx30cm TSKgel SWXL columns in series; two 10µm, 7.5mmIDx60cm TSKgel SW columns in series

Elution:

Globular proteins: 0.3M NaCl in 0.1M (0.05M for SWXL columns) phosphate buffer, pH 7
Dextrans and polyethylene glycols and oxides (PEOs): distilled water

 


Return to Size Exclusion main page

Application Overview -- Gel Filtration Chromatography with TSKgel columns

Tosoh Bioscience Japan, Shiba, Minato-Ku, Tokyo 105-8623, Japan TEL:+81-3-5427-5118 FAX:+81-3-5427-5198 Email:hlc@tosoh.co.jp