TSKgel SWXL columns, introduced in 1987, represent the second generation of high performance Gel Filtration columns that have become synonymous with analyzing protein molecular weights in the emerging field of biotechnology.
Particle Characteristics TSKgel SWXL-type columns contain smaller particles than SW-type columns; 5 and 8μm versus 10 and 13μm. As the SW-type columns, TSKgel SWXL columns feature highly porous silica particles, the surface of which has been shielded from interacting with proteins by derivatization with ligands containing diol functional groups. TSKgel SWXL-type columns stand out from other silica- or polymer-based high performance size exclusion columns by virtue of their large pore volumes. These columns are commonly used in the quality control of monoclonal antibodies and other biopharma products.
|
Calibration Curves The Figure below shows the calibration curves for globular proteins, polyethylene oxides and dextrans for each of the three TSKgel SWXL-type columns. The following Table summarizes the molecular weight ranges for G2000SWXL, G3000SWXL and G4000SWXL columns for globular proteins, polyethylene oxides and dextrans that can, at least partially, access the pores. For comparative purposes, similar data is provided as well for TSKgel SW and SuperSW-type columns.
|
Protein Calibration curves for TSKgel SWXL columns

| |
Properties and separation ranges for TSKgel SW-type packings (Molecular Weight in Da)
|
TSKgel packing |
Particle size (µm) |
Pore size (nm) |
Globular proteins |
Dextrans |
Polyethylene glycols and oxides |
|
SuperSW2000 |
4 |
12.5 |
5,000 - 150,000 |
|
|
|
G2000SWXL BioAssist G2SWXL |
5 |
12.5 |
5,000 - 150,000 |
|
|
|
G2000SW |
10, 13 |
12.5 |
5,000 - 100,000 |
|
|
|
|
|
|
|
|
|
|
SuperSW3000 |
4 |
25 |
10,000 - 500,000 |
|
|
|
G3000SWXL BioAssist G3SWxl |
5 |
25 |
10,000 - 500,000 |
|
|
|
G3000SW |
10 |
25 |
10,000 - 500,000 |
|
|
| |
|
|
|
|
|
|
G4000SWXL BioAssist G4SWXL |
8 |
45 |
20,000 -7,000,000 |
|
|
|
G4000SW |
13, 17 |
45 |
20,000 -7,000,000 |
|
| |
Data generated using the following conditions:
|
Columns: |
Two 4µm, 4.6mmIDx30cm TSKgel SuperSW columns in series; two 5µm, 7.8mmIDx30cm TSKgel SWXL columns in series; two 10µm, 7.5mmIDx60cm TSKgel SW columns in series |
|
Elution: |
Globular proteins: 0.3M NaCl in 0.1M (0.05M for SWXL columns) phosphate buffer, pH 7 Dextrans and polyethylene glycols and oxides (PEOs): distilled water |
|
 |
Return to Size Exclusion main page
Application Overview -- Gel Filtration Chromatography with TSKgel columns |